Machine technical sheet


Immagine rappresentane il macchinario
Brand: MEDELA
Model: Bilibed 0383012
Type: Electromedical equipment: neonatal phototherapy
Power supply: Electrical (220V-380V)
Reference standard: CEI EN 60601-2-50
Classification according to standard IEC-EN 62471 / IEC-EN 60825-1: Non disponibile
Category according to standard EN12198-1 Non disponibile

SOURCE TECHNICAL SHEET
Type: lampada a scarica di GAS
Emission: Continua
N. of lighting bodies: 1
Brand: OSRAM
Model: Osram Dulux L 18W/71
Classification according to standard IEC-EN 62471 / IEC-EN 60825-1: Non disponibile
Reference standard: CEI EN 60335-2-27/A2

RELEVANT EMISSION TYPES  
                                       
            780 ÷ 3000 (IRA e IRB)    
            780 ÷ 1400 (IRA)                    
        380 ÷ 1400
(Visible and IRA)
                   
      315 ÷ 400
(UVA)
                             
    300 ÷ 700
(Blue light)
                             
  180 ÷ 400
(UVA, UVB e UVC)
                             
⇐ BAND ⇒    
Main protection measures to implement for safety purposes:
  1. Do not look directly into the light beam

Measurements in the field (Click to view the measurements in the field)

Reference person:
AUSL 7 Siena - Servizio Prevenzione Protezione Rischi Daniele Cavarra ([email protected])

Rif.D.lgs 81/08 SPECTRAL RANGE TARGET ORGANS RESULT Val.
a ES (Ultraviolet) Eyes and skin Irrilevante
b EUVA (Ultraviolet A) Eyes da 20% ÷ 50% del limite 0.086 W m-2
c,d LB (Blue light, extended source) Eyes da 50% ÷ 80% del limite 70.8 W m-2sr-1
e,f EB (Blue light, small source) Eyes Non applicabile
  LAFA (Blue light, extended source, aphakic individuals) Eyes da 50% ÷ 80% del limite 77.6 W m-2sr-1
  EAFA (Blue light, small source,aphakic individuals) Eyes Non applicabile
g,h,i L R (Visible and Infrared A) Eyes Irrilevante 720 W m-2sr-1
j,k,l L R (Infrared A) Eyes Irrilevante 0.59 W m-2sr-1
m,n EIR (Infrared A + Infrared B) Eyes Irrilevante 0.19W m-2
o Eskin (Visible + Infrared A + Infrared B) Skin Irrilevante 19 W m-2
SENSOR-SOURCE DISTANCE
20
cm
Reference person:
AUSL 7 Siena - Servizio Prevenzione Protezione Rischi Daniele Cavarra ([email protected])

Rif.D.lgs 81/08 SPECTRAL RANGE TARGET ORGANS RESULT Val.
a ES (Ultraviolet) Eyes and skin Irrilevante
b EUVA (Ultraviolet A) Eyes Irrilevante 0.03 W m-2
c,d LB (Blue light, extended source) Eyes da 50% ÷ 80% del limite 76.5 W m-2sr-1
e,f EB (Blue light, small source) Eyes Non applicabile
  LAFA (Blue light, extended source, aphakic individuals) Eyes Maggiore del VLE 83.8 W m-2sr-1
  EAFA (Blue light, small source,aphakic individuals) Eyes Non applicabile
g,h,i L R (Visible and Infrared A) Eyes Irrilevante 780 W m-2sr-1
j,k,l L R (Infrared A) Eyes Irrilevante 70 W m-2sr-1
m,n EIR (Infrared A + Infrared B) Eyes Irrilevante 0.08W m-2
o Eskin (Visible + Infrared A + Infrared B) Skin Irrilevante 6 W m-2
SENSOR-SOURCE DISTANCE
40
cm
logo regione toscana
Regione Toscana
Diritti Valori Innovazione Sostenibilità
collegamento sito usl 7 siena Servizio Sanitario
della
Toscana