Machine technical sheet


Immagine rappresentane il macchinario
Brand: FASTER
Model: BHA48
Type: Laboratory sterile hood
Power supply: Electrical (220V-380V)
Reference standard: CEI EN 62471
Classification according to standard IEC-EN 62471 / IEC-EN 60825-1: Non disponibile
Category according to standard EN12198-1 Non disponibile

SOURCE TECHNICAL SHEET
Type: lampada a scarica di GAS
Emission: Continua
N. of lighting bodies: 1
Brand: Philips
Model: TUV30W G30T8
Classification according to standard IEC-EN 62471 / IEC-EN 60825-1: Non disponibile
Reference standard: CEI EN 62471

RELEVANT EMISSION TYPES  
                                       
            780 ÷ 3000 (IRA e IRB)    
            780 ÷ 1400 (IRA)                    
        380 ÷ 1400
(Visible and IRA)
                   
      315 ÷ 400
(UVA)
                             
    300 ÷ 700
(Blue light)
                             
  180 ÷ 400
(UVA, UVB e UVC)
                             
⇐ BAND ⇒    
Main protection measures to implement for safety purposes:
  1. Harmless when the hood is closed
  2. Protect the skin and the eyes if the shield is not present

Measurements in the field (Click to view the measurements in the field)

Reference person:
AUSL 7 Siena - Laboratorio Agenti Fisici Andrea Bogi ([email protected])

Rif.D.lgs 81/08 SPECTRAL RANGE TARGET ORGANS RESULT Val.
a ES (Ultraviolet) Eyes and skin Maggiore del VLE 4.85 W m-2
b EUVA (Ultraviolet A) Eyes da 20% ÷ 50% del limite 0.079 W m-2
c,d LB (Blue light, extended source) Eyes Irrilevante
e,f EB (Blue light, small source) Eyes Non applicabile
  LAFA (Blue light, extended source, aphakic individuals) Eyes Irrilevante
  EAFA (Blue light, small source,aphakic individuals) Eyes Non applicabile
g,h,i L R (Visible and Infrared A) Eyes Irrilevante
j,k,l L R (Infrared A) Eyes Irrilevante
m,n EIR (Infrared A + Infrared B) Eyes Irrilevante
o Eskin (Visible + Infrared A + Infrared B) Skin Irrilevante
SENSOR-SOURCE DISTANCE
20
cm
MAXIMUM EXPOSURE TIME
6
seconds